Abstract
The amount, identity, and size distribution of particles in parenteral therapeutic protein formulations are of immense interest due to potential safety and efficacy-related implications. In this communication, we describe the use of a flow cytometer equipped with forward- and side-scattering as well as fluorescence detectors, to determine the number of subvisible particles in monoclonal antibody formulations. The method appears to detect particles of size 1 μ and larger, requiring relatively small sample volumes to estimate subvisible particle counts. Additionally, it facilitates differentiation of proteinaceous particles after staining with a fluorescent hydrophobic dye. The method is expected to be particularly well suited for pharmaceutical development, because it provides increased throughput due to the use of a 96-well autosampler.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.