Abstract

A constant pressure gas flowmeter using a directly driven diaphragm bellows as a volume displacer was studied. This flowmeter is perfectly vacuum-sealed, does not contain elastomers and liquids that would prevent outgassing at elevated temperatures and can achieve a smaller ratio of the final volume to the displaced volume than when bellows are used, so that the uncertainty in generating small flow rates can be small. As, when a bellows is used in the volume displacer, the displaced volume cannot be calculated from the geometric dimensions and its dependence on the displacement is non-linear, a sensitive method had to be developed for measuring this dependence. The uncertainty in measuring the displaced volume by this method and the achieved uncertainty of the flowmeter were determined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.