Abstract
To determine the absorbed dose to the skin of patients undergoing radiotherapy with megavoltage radiations, it is necessary for the dose to be measured at shallow depths. This report assesses the suitability of carbon-loaded thermoluminescent dosimeters (manufactured by Vinten Instruments Limited) for this purpose. Measurements have been made of the effective measurement depth of the dosimeters, the linearity with dose, and the variation of sensitivity within a batch. The variation of surface dose with incident angle of radiation for 5-MV x rays has been measured using these disks; the results are compared with results obtained using a thin-windowed ionization chamber.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.