Abstract

Kvarstein, B. The Use of an Electronic Particle Counter (Celloscope 101) for Counting Leucocytes. Scand. J. din. Lab. Invest. 19, 196-202 1967. A method for counting leucocytes with an electronic particle counter using a Cetrimide solution as hemolyzing diluent was examined.The electronic particle counter gives reliable results with a coefficient of variation of 4.4 per cent.The leucocyte counts can be done immediately, or at any time within 24 hours with the same result.Rise in temperature led to a slight fall in white cell counts.The procedure did not require siliconized glass tubes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.