Abstract
A low-cost ambulatory blood pressure measuring device has been developed from a commercially available stationary apparatus. The device, which has been compared for accuracy with a mercury manometer, has functioned satisfactorily as an ambulatory monitor for 4 years. To minimize errors, blood pressure readings were made in the sitting position at half-hourly intervals over the waking day. The utility of the device in clinical trials has been investigated. On repeated readings no placebo effect on blood pressure was detectable and the mean difference between two readings in 42 subjects was 1.9/-0.33 mmHg. The standard deviation of the difference were 8.1/5.6 mmHg. This should make it possible to detect differences of 8/5 mmHg between two treatments in about 16 subjects. There was no detectable tendency for blood pressure to change during the day but the variability between readings was substantially increased if the observation periods were reduced to 4 h.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.