Abstract

To estimate the radiation dose reduction to the thyroid for an anterior oblique occlusal view from the use of a thyroid shield, compare this with the variation in thyroid dose resulting from differences in examination positioning and discuss the additional considerations associated with the use of a thyroid shield before making a recommendation on their routine use for this examination. Doses to the oral mucosa, the salivary glands, the thyroid, the extrathoracic airways, the oesophagus and the lungs were directly measured for anterior oblique occlusal X-rays of a Rando phantom with and without a thyroid shield using strips of calibrated XRQA Gafchromic film. The examination was also simulated using Monte Carlo software for the without thyroid shield case for a comparison of the dose and to evaluate the dosimetric effect of suboptimal examination positioning. A 36% reduction in thyroid dose was measured as a result of thyroid shield use; the effective dose reduction is of the order of 22%. Suboptimal positioning was found to increase thyroid dose by a far more significant amount. Despite the reduction in thyroid dose, cost-benefit considerations mean that the purchase of a thyroid shield is only recommended where a very high number of anterior oblique occlusal views are undertaken. Optimization efforts for this examination are better focussed on training in examination positioning.