Abstract

A retrospective study was performed to evaluate the cement mantle in two groups of patients treated with the acetabular components of cemented Exeter total hip arthroplasties (THAs). Two groups of 20 patients were compared: Group 1 received non-flanged acetabular cemented cups (Contemporary, Stryker) and Group 2 received flanged acetabular cemented cups (X3 Rim Fit, Stryker). Cups in Group 2 were implanted after using a rim cutter device. Group 2 showed better penetration of cement in zone 1 (10.76mm compared with 2.93mm; p=0.008) and a thicker cement mantle in zone 1 (3.57mm compared with 2.89mm; p=0.04). More cups in Group 2 had a cement mantle thickness less than 3mm (30% in Group 1 compared with 70% in Group 2; p=0.0039). No other radiological differences were observed. These results favor the use of a rim cutter device and flanged cup to improve the cement mantle for the acetabular components of cemented Exeter THAs. However, the improvements were less than expected. In view of the results of previous studies, further research is therefore needed to assess the value of this approach in improving the acetabular cement mantle.