Abstract

A method for determining platinum, rhodium, palladium and gold simultaneously on a direct-reading spectrometer with a non-absorbing reference line is described. The concentration ranges of the working curves were as follows: platinum, rhodium and palladium 0 to 20 p.p.m. and gold 0 to 5 p.p.m. The detection limit for rhodium and palladium was 0·1 p.p.m., for platinum 0·5 and for gold 0·01 p.p.m. The reproducibility of the results was satisfactory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call