Abstract

The catalytic abilities of gold electrode were tested for the quantitative determination of amphetamine (A) and 3,4-methylenedioxy-N-methylamphetamine (MDMA) standards by their oxidation using cyclic voltammetry (CV). The value of the oxidative currents of A and MDMA standards at 0.80 V vs. SCE in 0.05 M NaHCO3 at the scan rate of 50 mV/s is linear function of concentration in range of 110.9-258.9 mM and 38.7-229.2 mM, respectively. Square wave voltammetry (SWV) revealed linear increase of current with concentration of MDMA (range 30.9-91.6 mM) and thus quantitative determination of amphetamine derivates. SWV analysis is successfully performed in spiked urine samples as well. A and MDMA in the presence of sucrose and as a content in illegally produced tablets were also analyzed. The voltammetric determination of A and MDMA derivatives using CV and SWV at gold electrode is a rapid, selective and simple procedure and its accuracy was confirmed with reference method, high performance liquid chromatography (HPLC). The spiked urine samples analysis offers additional possibility for the rapid detection of A and MDMA in human urine.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call