Abstract

The use of the Association of Anaesthetists of Great Britain and Ireland checklist for anaesthetic machines, based on an oxygen analyser, was surveyed over a 5-week period in a teaching hospital. Fifty-five completed checklists were analysed; no problems developed during anaesthesia which were missed by the checklist. The mean time taken to complete the checklist for one machine was 8.9 min with a range of 5 to 19 min; for two consecutive machines it was 18.25 min with a range of 10 to 30 min. The most frequent faults detected were the poor reliability of some oxygen analysers, absent ventilator disconnection alarms, and absent oxygen supply failure alarms on some older machines. Faults were found in 60% of the machines checked; 18% of these were deemed to be serious.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.