Abstract
We describe the upgrade and performance of the high-purity germanium counting facility Gator, which is dedicated to low-background γ-ray spectrometry. Gator is operated at the Gran Sasso Underground Laboratory in Italy, at an average depth of 3600 meter water equivalent, and employed for material screening and selection in ultra-low background, rare-event search experiments in astroparticle physics. The detector is equipped with a passive shield made of layers of copper, lead and polyethylene, and the sample cavity is purged with gaseous nitrogen maintained at positive pressure for radon suppression. After upgrading its enclosure, the background rate is (82.0 ± 0.7) counts/(kg·day) in the energy region 100 keV to 2700 keV, a 20% reduction compared to the previously reported rate. We show the stability of various operation parameters as a function of time. We also summarize the sample analysis procedure, and demonstrate Gator's sensitivity by examining one material sample, a candidate photosensor for the DARWIN experiment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.