Abstract

In 2017, we published the first measurement result with the NIM-2 joule balance and the relative standard uncertainty was 2.4 × 10−7. Since then, a series of improvements have been carried out to upgrade the NIM-2 joule balance towards the 10–8 level relative uncertainty. The main improvements include a permanent magnet, as well as a two-layer magnetic shield to decrease the main uncertainty contribution from the external magnetic field, a parasitic displacement measurement system and a new suspension for improving alignments, and a new mass translation stage constructed with all non-magnetic materials. In September 2019, when the NIM-2 joule balance was used to calibrate a 1 kg mass, the type A relative standard uncertainty had been reduced to 3 × 10−8(k= 1) and the combined relative standard uncertainty was less than 1 × 10−7(k= 1). In October and November 2019, a 1 kg Pt/Ir mass and a 1 kg stainless steel mass were measured with the NIM-2 joule balance, and the relative standard uncertainties were 5.2 × 10−8 and 6.5 × 10−8, respectively. The relative difference between the measured values and the calibration values obtained from the mass group were 4 × 10−9 and 3 × 10−8, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.