Abstract
Filtering and absorbing products in the form of a plate and working under conditions of convective blowing with cleared air residual life indicator uncertainties of measurement analysis is carried out. It is shown that for 0.8 mm thick KO2-based sorbent plates at a relative humidity of 60…90 % and a CO2 concentration of 1…4 %, the main factor of the uncertainty of indirect measurement of the residual life indicator is the Biot parameter characterizing the conditions of heat transfer from the plate surface. It has been experimentally proven that for Biot parameter in the range of 0–0.08, this factor contributes up to 30 % in the total uncertainty.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.