Abstract

A description is given of the UK primary standard graphite calorimeter system. The calorimeter measures absorbed dose to graphite for photon radiations from to 19 MV x-rays, and is the basis of the NPL therapy-level absorbed dose to water calibration service. Absorbed dose to graphite from the photon calorimeter has been compared with three other standards: an ionization chamber and cavity theory, for gamma radiation; the NPL electron calorimeter, for 12 - 14 MeV electron beams; and the BIPM absorbed dose standard. The three standards agreed within 0.5% which is similar to the measurement uncertainties.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.