Abstract

A new device, the TRIple PLunger for EXotic beams (TRIPLEX), has been developed for lifetime measurement studies with rare isotope beams. This plunger device holds up to three metal foils in the beam path and facilitates the recoil distance Doppler-shift technique to measure lifetimes of nuclear excited states in the range of 1ps to 1ns. The unique design allows independent movement of the target and the second degrader with respect to a fixed first degrader in between, enabling advanced experimental approaches, such as the differential recoil distance method and the double recoil distance method. The design and control of the device are presented in this paper, together with simulated performances of the new applications. As an example of actual experiments, results from the lifetime measurement of the neutron-rich 17C isotope performed at the National Superconducting Cyclotron Laboratory are shown.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.