Abstract

The increasing use of more rigid internal fixation constructs for spinal fractures, especially in association with spinal cord injury, has led surgeons to combine sublaminar segmental wiring with Harrington instrumentation systems. Two clinical cases whose neurologic condition deteriorated postoperatively were shown to have sustained direct cord injury by the combination of Harrington compression rods with segmental sublaminar wiring. Myelographic and surgical evidence of hook protrusion into the spinal canal with direct cord injury is presented. Laboratory spine simulations duplicating the clinical situation did demonstrate that sublaminar wiring of the Harrington compression rod system caused the standard hooks to protrude dangerously into the spinal canal. Caution should be exercised not to combine Harrington compression rods with segmental sublaminar wiring.