Abstract

Comparative measurements were made of depth-dose profiles in an Alderson Rando phantom and a similarly shaped phantom of water which closely simulates the attenuation properties of human muscle. Measurements were carried out at three different kV settings of the X-ray generator, covering the lower-energy range of normal diagnostic practice over which attenuation differences are likely to be most important. The results show that whilst the Alderson phantom can undoubtedly be extremely useful in medical dosimetry, corrections to the dose distributions observed for diagnostic qualities of radiation may sometimes be necessary to allow for departures from tissue-equivalence.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.