Abstract

Implantable antitachycardia devices rely upon schemes for detecting cardiac arrhythmias which utilize rate and its variations; yet rate parameters often identify nonpathologic tachycardias as potentially dangerous and deliver unwarranted therapy. I have developed a predictive filter based upon the time-sequenced adaptive algorithm to be used as a supplement to rate criteria for detecting and identifying serious arrhythmias. The method does not require a fixed template and is independent of a priori patient information. The algorithm also provides arrhythmia diagnosis immediately at the change in rhythm. Algorithmic parameters were determined based upon a training set of patient data, and performance of the technique was evaluated with a completely new test set of 20 arrhythmia passages. The new algorithm yielded a sensitivity and specificity for ventricular tachycardia of 91% and 82% and for ventricular fibrillation of 71% and 93%. Correlation waveform analysis was used to diagnose the same test set of arrhythmias. It yielded a sensitivity and specificity for ventricular tachycardia of 100% and 67% and for ventricular fibrillation of 50% and 100%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.