Abstract

A platinum resistance thermemeter will be used to determine the vertical temperature profile of the Titan lower atmosphere. The sensor is part of the HASI (Huygens Atmospheric Structure Instrument) which will be located on the Huygens probe for the Cassini mission. The design and manufacture of a laboratory prototype have made it possible to verify calibration and test procedures appropriate to the flight units. The sensor has been selected and evaluated with great care in order to ensure performance within the mission requirements which demand accuracy and resolution of 0.5 and 0.02 K, respectively. The calibration and test results confirm performance within the design goals. Further results, dealing with mechanical characterizations aimed at verifying the sensor's capability to withstand launch, entry and descent conditions, are also presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.