Abstract
The femoral neck system (FNS) was recently introduced for treatment of femoral neck fractures. A novel feature is an anti-rotation screw with a diverging design from the neck bolt to provide rotational stability. Using the 4Nm torque limiting screwdriver might cause the incomplete insertion of the screw specifically in dense or sclerotic bone. A nontorque limiting manual screwdriver should be used to complete the insertion and locking of the screw to prevent this error. We present series of seven cases with incomplete insertion of the anti-rotation screw in patient with femoral neck fractures treated with the FNS.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European journal of orthopaedic surgery & traumatology : orthopedie traumatologie
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.