Abstract
Study of the time characteristics of photomultiplier tube with ultra-fast time characteristics has an important guiding role for the further development of ultra-fast time response microchannel plate photomultiplier tube (FPMT). Based on the VME test system in high-energy physics and picosecond laser with single-photon pulse mode, this manuscript designs a device to test the FPMT with 25 ps system error. The time characteristics of various FPMTs were tested by optimizing the FPMT signal readout anode, the voltage divider structure and voltage division ratio. The intrinsic time lower limit value of the FPMT in the non-single-photon working mode, is proposed to compare and analyze the time resolution of different FPMTs in different working states. After completing various optimized readout anode structural design for the FPMTs, it can be find that the best FPMT prototype in our Lab has the best intrinsic time resolution lower limit of 30 ps.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.