Abstract

The linearity and photon detection efficiency (PDE) are important parameters of photomultiplier tube (PMT), which need to be precisely measured with suitable techniques. To search good methods for linearity and PDE study of newly developed 20-inch PMT. In this paper, we setup a testing system to use the average photoelectrons (P.E.) in pulse mode for linearity measurement with the result corrected for strong light. For the PDE measurement, we developed a relative method with a reference PMT whose efficiency is known to compare the measured PDE with both average charge and hit numbers. The measurements of PMT shows only 5% nonlinearity within 1000 P.E. and good PDE, the result of linearity’s corrected definition is closer to the real status of PMT. The results show good charge and linearity response for newly developed 20-inch MCP PMT and dynode PMT. The testing techniques discussed in this paper are suitable for the study of PDE and linearity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.