Abstract

Several different methods of file sterilization were analyzed to determine the best method of providing complete file sterility, including the metal shaft and plastic handle. Six test groups of 15 files were studied using Bacillus stearothermophilus as the test organism. Groups were "sterilized" by glutaraldehyde immersion, steam autoclaving, and various techniques of salt sterilization. Only proper steam autoclaving reliably produced completely sterile instruments. Salt sterilization and glutaraldehyde solutions may not be adequate sterilization methods for endodontic hand files and should not be relied on to provide completely sterile instruments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.