Abstract

Accuracy of a stereotactic operating microscope, by which imaging data may be superimposed on the operative field without a stereotactic frame, has been most limited by the resolution of imaging information. Using newer algorithms and pilot pole calibration of the digitizer, an error in registration of 2 mm and in contour display of 3 mm has been demonstrated. Greatest utility of the system clinically has been in providing navigational guidance to small lesions undergoing resection.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.