Abstract

The spindle pump is a combined working nonpulsatile blood pump, i.e., a centrifugally propelling device. This special concept was chosen to tackle the main problems of nonpulsatile pumps, such as thromboembolic complications, sealing difficulties, and traumatic hemolysis. The first two of these problems were relatively simple to solve compared with the third problem--traumatic hemolysis. Various modifications and constructive steps of the spindle pump were necessary to keep stress on blood components within adequate limits. The actual prototype--tested in three acute experiments and eight experiments with a longer pumping duration (the longest lasting 63 hours)--is described in extenso, and its development is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.