Abstract

Using a modeled look-up chart developed in this work, we show that accurate photometric measurements of characteristics like illuminance or luminous intensity of test LED sources can be measured in one step. This is a simple broadband measurement that may substitute for complicated and costly spectral measurements currently in place. To develop the look-up chart, the typical minimum and maximum of the Spectral Mismatch Correction Factor (SMCF) for a given photometers was estimated in relation to catalog parameters of the LEDs such as correlated color temperature (CCT) and melanopic daylight efficacy ratio (mDER). This research was based on the unique and large dataset of real photometers spectral response srel(λ) collected and measured at accredited laboratories located at America, Asia and Europe and modeled LED's spectral power distribution (SPD) data. Independent look-up tables were developed for color-mixed LEDs (cm-LEDs) and white phosphor-converted LEDs (pc-LEDs), two common LED types.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.