Abstract
This paper examines the effects of optically stimulated luminescence (OSL) components, other than that usually termed the fast component, on dose determination by the single-aliquot regenerative-dose (SAR) protocol. Results are presented for “dose recovery tests”, in which a known laboratory dose, delivered after optical bleaching at room temperature, is measured using the SAR protocol. Data obtained using either the initial OSL signal or the fast component, derived by curve fitting, are compared. Dose recovery tests are also carried out when an additional step is added to the SAR protocol, aimed at reducing recuperation, i.e. the residual signal observed in a SAR cycle when no regenerative dose is applied. The results for quartz from various sources indicate much improved dose recovery when only a well-separated fast component is analysed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.