Abstract

PurposeHeat and moisture exchangers (HMEs) and heated humidifiers (HHs) may be used together inadvertently. Such an incident occurred at our institution resulting in airway occlusion. Material and methodsA bench-top study was conducted to compare the incidence of airway occlusion when using (1) no airway humidification, (2) HME alone, (3) HH alone, and (4) both HME and HH in combination as part of a standard breathing circuit. ResultsThe simultaneous use of a HME and a HH was associated with a reduction in tidal volume (no airway humidification, P ≤ .05; HME alone, P ≤ .01; and HH alone, P ≤ .01) and an increased incidence of airway occlusion (no airway humidification, 0/7; HME alone, 0/7; HH alone, 0/7; and HME and HH in combination, 7/7; P < .0001). ConclusionsThe use of a HME and a HH in combination is likely to result in airway occlusion. Precautions should be taken to ensure that both systems are not used together in clinical practice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.