Abstract

Tritium activity in water samples has been measured by means of a proportional counter. The energy interval was set between 1.1 keV and 9.7 keV to obtain optimal efficiency (the best figure of merit). The background count rate in the energy interval above 20 keV was measured simultaneously. A very good correlation between background count rates in both energy intervals has been observed. The counting gas quality has been checked by measuring the 55Fe spectral line (5.89 keV). A good correlation between the count rate above 20 keV and the spectral resolution has been obtained enabling gas quality control during the measurement. A constant monitoring of the count rate above 20 keV has revealed radon contamination of several groundwater samples. Repeated measurement after 15 days storage gave more accurate tritium activity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.