Preconcentration method with collection of metal complexes on a chitin has been applied to the spectrophotometric determination of chromium(VI) in water. The chromium(VI) is collected as its 1,5-diphenylcarbazide(DPC) complex on a column of chitin in the presence of dodecyl sulfate as counter-ion. The Cr-DPC complex retained on the chitin is eluted with a methanol-1 M acetic acid mixture (7:3, v/v), and the absorbance of the eluent is measured at 541 nm. Beer’s law is obeyed over the concentration range of 0.05–0.6 μg of chromium(VI) in 1 ml of the eluent. The apparent molar absorptivity is 3.5×10 4 dm 3 mol −1 cm −1. The tolerance limits for Fe(III) is low, i.e. ten times that of chromium(VI), but some metal ions and common inorganic anions do not interfere in concentration range of 100–10 000 times that of chromium(VI). The present method can be applied to the determination of chromium(VI) in natural water samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.