Abstract

Five fresh cadaver upper extremities were studied by use of a static positioning frame, pressure-sensitive film, and a microcomputer-based videodigitizing system, to assess the load-bearing characteristics of a scaphoid silicone implant within the radioulnarcarpal joint. Specimens were studied in their "normal" condition, after resection of the scaphoid, after placement of a scaphoid implant, and with a scaphoid implant and a simulated capitate-lunate-triquetrum-hamate fusion. The scaphoid silicone implant bore significant, although less, load than the normal scaphoid. Decreasing the size of the scaphoid implant decreased the load born by the implant. Decreased load through the scaphoid implant was compensated by the lunate. The addition of a limited carpal fusion did not significantly decrease the load born by a scaphoid implant. Therefore, the silicone scaphoid implant is a load-bearing implant even when undersized or placed in association with a limited carpal fusion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.