Abstract

The aim of the present study is to investigate bone tissue reactions to various surface oxide properties, in particular to different surface oxide chemistry of oxidized titanium implants (grade 1). One control and three test screw-shaped implant groups were prepared. Controls were turned implants. Test implants, i.e. S implants, P implants and Ca implants were by the micro-arc oxidation (MAO) method. The surface characterizations were performed with X-ray photoelectron spectroscopy, Auger electron spectroscopy, scanning electron microscopy, X-ray diffractometry and TopScan 3D ®. Eighty implants were inserted in the femora and tibiae of ten mature New Zealand white rabbits for 6 weeks. The removal torque values (RTQ) showed significant differences between S implants and controls ( p=0.022), Ca implants and controls ( p=0.0001), Ca implants and P implants ( p=0.005) but did not show significant differences between the others ( p>0.05). In addition, the bone to metal contact (BMC) around the entire implants demonstrated 186% increase in S implants, 232% increase in P implants and 272% increase in Ca implants when compared to the paired control groups. Based on the comparative analysis of the surface characteristics resulting different bone responses between all groups, it was concluded that surface chemistry and topography separately or together play important roles in the bone response to the oxidized implants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.