Abstract

Precise calibration of the indenter shape is an important procedure in nanoindentation analysis since the indenter geometry enters directly into the most common methods of data analysis in this type of testing. Not only is the geometry required to be known with some precision, but also the sharpness of the tip, especially in the case of pyramidal indenters, is important for the use of indenters for testing hardness in thin film specimens—the most common application of nanoindentation. In this paper, a method of determining the area function and tip radius for a Berkovich indenter is described. It is shown that the tip radius estimated from the area function data is in reasonable agreement with a direct measurement using a calibrated atomic force microscope. It is shown that subjective decisions about tip radius may lead to unjustified rejection of a tip for hardness measurement. A new criterion for tip quality is presented in terms of tip radius and specimen material properties.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.