Abstract

AbstractBackgroundDue to the increased degree of modulation and complexity of volumetric-modulated arc therapy (VMAT) plans, it is necessary to have a pre-treatment patient-specific quality assurance (QA) programme. The gamma index is commonly used to quantitatively compare two dose distributions. In this study we investigated the sensitivity of single- and multi-gamma criteria techniques to detect multileaf collimator (MLC) positioning errors using the Varian TrueBeam Electronic Portal Imaging DeviceTM (EPID) dosimetry and the ArcCHECKTM device.Materials and methodsAll active MLC positions of seven intact prostate patients VMAT plans were randomly changed with a mean value of 0.25, 0.5, 1 and 2 mm and a standard deviation of 0.1 mm on 25, 50, 75 and 100% of the control points. The change in gamma passing rates of six gamma criteria of 3%/3 mm, 3%/2 mm, 3%/1 mm, 2%/2 mm, 2%/1 mm and 1%/1 mm were analysed individually (single-gamma criterion) and as a group (multi-gamma criteria) as a function of the simulated errors. We used the improved and global gamma calculation algorithms with a low dose threshold of 10% in the EPID and ArcCHECK software, respectively. The changes in the planning target volume dose distributions and the organs at risk due to the MLC positioning errors were also studied.ResultsWhen 25, 50, 75 and 100% of the control points were modified by the introduction of the simulated errors, the smallest detectable errors with the EPID were 2, 1, 0.5 and 0.5 mm, respectively, using the multi-gamma criteria technique. Similarly for the single-gamma criteria technique errors as small as 2, 1, 1 and 1 mm applied to 25, 50, 75 and 100% of the control points, respectively, were detectable using a 2%/2 mm criterion. However, the smallest detectable errors with the ArcCHECK when using the multi-gamma criteria technique were 2, 2 and 1 mm when MLC errors were applied on 50, 75 and 100% of the control points. When only 25% of the control points were affected the ArcCHECK were unable to detect any of the errors applied. No noticeable difference was observed in the sensitivity using the single- or the multi-gamma criteria techniques with the ArcCHECK.ConclusionThe Varian TrueBeam EPID dosimetry shows a higher sensitivity in detecting MLC positioning errors compared with the ArcCHECK regardless of using the single- or the multi-gamma criteria techniques. Higher sensitivity was observed using the multi-gamma criteria technique compared with the single-criterion technique when using the EPID.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.