Abstract

A simple method for measuring the second order diffraction efficiency in the vacuum ultraviolet (VUV) is described. Spectral reflectance of a mirror will be influenced by the second diffraction order, measured in a spectrophotometer system without filters. The second order diffraction efficiency can be calculated from the different spectral reflectance values. A deuterium lamp and a scintillated photomultiplier are used in the measurement system. The second order diffraction efficiency can be determined from 120 nm to 165 nm. The result of 0.00579+/-11.9% is obtained at 161 nm.