Abstract
Nitric oxide (NO) inhalation and high frequency oscillatory ventilation (HFOV) has been indicated in infants with severe respiratory failure. The purpose of the present study was to evaluate the safety of an NO inhalation system with HFOV in terms of nitrogen dioxide (NO2) production. The NO inhalation system consisted of a high frequency oscillatory ventilator, a neonatal circuit and a test lung. The NO concentration was changed from 0 to 19 p.p.m. At each level of NO, the oxygen (O2) concentration was changed from 21 to 100%. The NO and NO2 concentrations were measured with a chemiluminescence analyzer using a molybdenum converter. The NO2 concentration was increased when either the O2 or the NO concentration was increased. The interposition of the endotracheal tubes increased NO2 concentrations at 4 p.p.m. NO. The high stroke volume and high mean airway pressure produced a significant increase in NO2 production at 4 p.p.m. NO. The increase in NO2 production was prevented by placing a one-way valve at the joint of the NO gas line to the inspired limb. It was concluded that the NO inhalation system with HFOV can be safely used when a one-way valve is placed at the joint of the NO gas line to the inspired limb and when inhaled NO is at a relatively low concentration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.