Abstract
A detailed derivation is presented of the formulas required to determine Ngas and Dmed in the AAPM TG-21 dosimetry protocol. This protocol specifies how to determine the absorbed dose in an electron or photon beam when using exposure or absorbed dose calibrated ion chambers. It is shown that the expression given in TG-21's recent letter of clarification is incorrect. Accounting for humidity correctly increases, by 0.4%, all absorbed dose determinations using an exposure calibrated ion chamber. Taking into account other correction factors in the equation for exposure could also have varying, but significant effects (possibly over 1%). These are the stem scatter correction, the axial nonuniformity correction and the electrode correction for electrodes made of different materials from the wall. Attention is drawn to differences in the definitions of the exposure and absorbed dose calibration factors, Nx and ND, respectively, as supplied by the NBS and the NRCC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.