Abstract
Cortical ERA was conducted in a consecutive series of 118 medicolegal cases (467 threshold determination in 209 ears). In organic cases, the distribution of subjective (manual pure tone audiometry)/objective (CERA) discrepancies was normal and centred on 0 dB, with only 3.2% of the discrepancies exceeding + 15 dB. Averaged over 3 or more test frequencies in one ear, only 4.4% of the discrepancies exceeded + 7.5 dB, as compared to 95.1% exceeding + 7.5 dB in those suspected of having a non-organic overlay. The pattern of non-organic overlay superimposed on noise-induced hearing loss is expected and demonstrated to produce a flattening of the dip in the audiogram. It is argued that such flattening is very rarely the end product of noise exposure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.