Abstract

The single scattering model for gamma backscatter density gauges has been extended to describe how the total detected count rate changes in response to localised density variations within the material. This extended model suggests there is a spatial region where density perturbations have a contradictory effect on the measured density value, an effect that has already been shown experimentally by previous workers. Here we compare their results with those predicted by application of the extended single scattering model. Since a complete description of their experimental apparatus was not available, only a crude fit could be achieved. Nevertheless, all the basic features of the data could be reproduced.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.