Abstract

In this paper, we describe how the resolution integral can be used as a tool for characterising the grey-scale imaging of diagnostic ultrasound scanners. The definitions of resolution integral, characteristic resolution and depth of field are discussed in relation to grey-scale imaging performance, together with a method of measuring these parameters using the Edinburgh Pipe Phantom. We show how the characteristic resolution and depth of field can be used to quantify the differences between transducers designed for different applications and how they are useful in identifying and quantifying changes in the performance of individual transducers.