Abstract

X-ray diffraction from LaB 6 standards document a precision of 478 ppm in lattice-parameter determinations for beamline 12.2.2 at Lawrence Berkeley National Laboratory′s Advanced Light Source, a facility for characterizing materials at high pressures and temperatures using laser- and resistance-heated diamond cells. Melting of Ni, Mo, Pt and W, resistively heated at 1 atm pressure in Ar, provides a validation of the beamline spectroradiometric system that is used to determine sample temperatures. The known melting temperatures, which range from 1665 to 3860 K for these metals, are all reproduced to within ±80 K.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.