Abstract

The resistive bolometer system has been successfully employed on experimental advanced superconducting tokamak for the first time to measure the radiated power of plasma. The bolometer detectors are based on 4 μm thick Pt absorbers deposited on 1.5 μm thick SiN membranes. The system consists of 3 cameras with a total of 48 channels. The detector and the system setup are described in detail. The detector calibration and typical measurement results are presented as well.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.