Abstract
Ninety urine samples were examined twice by 3 "observers" (two persons, using only visual observation, and one person using a spectrophotometric analyser) using multiple reagent teststrips. To determine reproducibility, inter- and intra-observer agreement were calculated and expressed as Cohen's kappa and as weighted kappa. The results show negligible intra-observer differences between the visual and spectrophotometric observation. The lack of agreement between inter- and intra-observer urinalysis results, using multiple reagent test strips was disappointing, considering the simplicity of the test procedure. Further improvement of reproducibility, e.g. by enhancing the discoloration of the test pads, is necessary. Reproducibility is not improved by using a spectrophotometric analyser instead of visual reading of the test strips.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.