Abstract

In an effort to replace the complement-dependent cytotoxicity test (CDC) with a more sensitive single technique we looked at flow cytometry as a possible replacement. The Flow PRA Bead technique (One Lambda) performed well in our laboratory. Although as expected this technique was more sensitive and specific than CDC, there remained 11 samples from eight patients which were flow negative, CDC positive. The results of various antibody identification tests on these samples prompted us to alter the positive selection criteria which we had been using on our routine screening with the Flow PRA Beads and persuaded us that the initial CDC result was correctly positive in nine of the 11 samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.