Abstract

BackgroundIn pursuit to improve soft tissue balancing in total knee arthroplasties (TKAs), a wireless device was introduced to assess femorotibial pressures. The aim of this study was to evaluate the reliability of this device. MethodsAfter 33 TKAs were balanced by conventional techniques, contact pressures were measured using a wireless sensor 3 times in a row; twice while the examiner was blinded for the result (n = 29); and once while the examiner was able to see the result as visual feedback (n = 32). Femorotibial pressures were measured in the medial and lateral compartments with the knee in 10°, 45°, and 90° of flexion (6 measurements per TKA). Furthermore, both the combined pressure and the difference in pressure between the compartments was calculated throughout the 3 positions (together another 6 measurements per TKA). ResultsThe intraclass correlation coefficient between the blind measurements was poor in 2 of the 12 (17%), moderate in 4 of 12 (33%), and good in 6 of 12 (50%) measurements. The intraclass correlation coefficient between the blind and observing measurement was poor in 2 of the 12 (17%), moderate in 6 of 12 (50%), and good in 4 of 12 (33%) measurements. Especially measurements in 10° of flexion are associated with poorer reliability. ConclusionThe wireless sensor has a moderate to good reliability in 83% of the measurements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.