Abstract

Physeal closure after slipped capital femoral epiphysis fixation can be difficult to assess on two-dimensional conventional radiographs. Radiostereometric analysis offers improved motion detection over conventional radiography, whereas the EOS biplanar imager provides a means for low radiation weight-bearing images. This phantom study assessed the reliability of measuring motion using radiostereometric analysis in the EOS using a slipped capital femoral epiphysis model. The accuracy and precision were better than 0.09±0.05 mm and 0.20°±0.36° when centered in the imaging space, were within the limits of clinical significance, and were not different from a standard uniplanar radiostereometric system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.