Abstract

BackgroundThe purpose of our cadaveric study was to determine the proximity of nail insertion and interlocking mechanisms in the Phantom® Lapidus Intramedullary Nail System to neurologic and tendinous structures in the foot. MethodsWe used 10 fresh-frozen human lower-extremity specimen cadavers. For each specimen, the Nail System was inserted as described in the published technique guide. We then performed dissection on the tibialis anterior tendon, extensor hallucis longus tendon, and medial dorsal cutaneous branch of the superficial peroneal nerve and we measured and averaged the distances from each of these structures from the nail. ResultsThe tibialis anterior tendon was in closest proximity to the insertion of the proximal medial interlock K-wire with an average distance of 0.4mm from the tendon. The extensor hallucis longus tendon was in closest proximity to nail insertion with an average distance of 1.2mm. The medial dorsal cutaneous branch of the superficial peroneal nerve was in closest proximity to the distal interlock K-wire with an average distance of 7.5mm. ConclusionsThe tibialis anterior tendon, extensor hallucis longus tendon, and the medial dorsal cutaneous branch of the superficial peroneal nerve are at risk with the insertion of the nail system. Blunt dissection should be performed using this system with a path to bone before instrumentation to reduce the risk of nerve and tendon injury in the foot.