Abstract

This study sought to evaluate the relationship between post-stent strut apposition and follow-up strut coverage using contour plot optical coherence tomographic analysis. Tracking the fate of interested regions of struts at different time points has not been investigated. Post-intervention and 6-month follow-up optical coherence tomographic evaluations were performed in 82 patients treated with biolimus- (n = 37) or sirolimus-eluting stents (n = 45). Post-stent apposition was classified as embedded, apposed, or malapposed. For volumetric stent evaluation, the post-intervention strut-artery distance and the neointimal thickness at follow-up were measured as a function of the circumferential arc length and longitudinal stent length. Computer-generated contour plots of the strut-artery distance and neointimal thickness were compared. The percentages of embedded and malapposed struts after intervention were 1.8% (Interquartile range [IQR]: 0.6% to 6.2%) and 2.3% (IQR: 0.5% to 5.2%), respectively. The percentages of uncovered and malapposed struts at 6 months were 16.0% (IQR: 7.4% to 33.3%) and 0% (IQR: 0% to 0.7%), respectively. The percentage of uncovered struts at 6 months varied significantly with post-stent strut apposition (0% [IQR: 0% to 11.4%] in embedded, 16.3% [IQR: 8.1% to 31.3%] in apposed, and 26.8% [IQR: 0% to 56.3%] in malapposed, p < 0.001 for all pairwise comparisons). In lesions without tissue prolapse, embedded struts were all covered (100% covered struts) compared with those with tissue prolapse (76.8% covered, p < 0.001). The optical coherence tomography-guided optimization of stent strut apposition enhances strut coverage at follow-up. This comprehensive method for evaluating strut apposition may provide more useful information to understanding the serial changes in strut coverage. (Neointimal Coverage After Implantation of Biolimus Eluting Stent With Biodegradable Polymer: Optical Coherence Tomographic Assessment According to the Treatment of Dyslipidemia and Hypertension and the Types of Implanted Drug-Eluting Stents; NCT01502904).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.