Abstract

The aim of this study was to determine whether resonance frequency analysis can be integrated into the routine clinical evaluation of the initial healing of dental implants. In addition, this study was designed to verify whether there was a correlation between implant stability quotient (ISQ) values, maximum insertion torque values, angular momentum and energy, and to evaluate the importance of different clinical factors in the determination of ISQ values and maximum insertion torque values at implant insertion. Two different implant designs of 81 dental implants in 41 patients were evaluated using ISQ values. Maximum insertion torque values were obtained during the placement procedure. Two new methods were used to calculate the angular momentum developed due to implant installation as well as the energy absorbed by the bone. A linear correlation between ISQ values and maximum insertion torque values at the initial implant surgery was found (P < 0·01). There was a correlation between ISQ values and angular momentum (P < 0·05), although ISQ values and energy did not show a significant linear correlation at the initial surgery (P > 0·05). There was a correlation between maximum insertion torque values, each part's angular momentum, and their energies during installation (P < 0·01). The sequence of the variables that influenced ISQ values was implant location, design, diameter, and gender of the patient. The results of this experiment suggest that both ISQ values and new methods to calculate angular momentum and energy can help to predict implant stability.