Abstract

From January 2005 to July 2008, a retrospective study was conducted at a single institution to investigate technical success of the use of a reentry device (Outback LTD reentry catheter) in aortoiliac and femoropopliteal artery recanalization in 34 patients (18 men; mean age +/- SD, 72 years +/- 11) in whom the conventional guide wires and catheters failed to reenter the true lumen. True lumen reentry was achieved in 87% (n = 23) and 91% (n = 11) of patients with femoropopliteal and aortoiliac occlusions, respectively. The overall technical success rate with the device was 88% (n = 34). The device success rate in Transatlantic Inter-Society Consensus II class D lesions was significantly lower than in lower lesion classes (71.4% vs 100%; P < .05). No procedure-related complications were encountered. In conclusion, the use of the reentry catheter enhances the likelihood of successful subintimal recanalization of chronic occlusions in femoropopliteal and aortoiliac arteries.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.